  • Non-invasive Treatment Efficacy Evaluation for HIFU Therapy Based on Magneto-Acousto-Electrical Tomography
    Non-invasive Treatment Efficacy Evaluation for HIFU Therapy Based on Magneto-Acousto-Electrical Tomography

    To ensure the therapeutic effect of HIFU therapy with minimized damage to surrounding tissues, a monitoring method for thermal ablation using the MAET technology is proposed based on the temperature-conductivity relation of tissues. With the distributions of acoustic pressure, temperature and electrical conductivity for a cylindrical model, real-time simulations of MAET signal demonstrated that two wave clusters can be generated by the sharp conductivity variation of HIFU ablation at >69 ˚C with a minimum axial interval of one wavelength. The favorable results provide a sensitive modality for non-invasive treatment efficacy evaluation during HIFU therapy and suggest potential applications in biomedical engineering.